A passive shimming method for Halbach magnet based on magnetic sheet arrays #Ustrumentation

Published: Wednesday, 02 November 2022 - 10:00 -0400

Author: Thorsten Maly

Wang, Ya, Yajie Xu, Feng Wang, Junhao Zhang, Bowen Peng, Peng Yu, Yingcong Yu, and Xiaodong Yang. “A Passive Shimming Method for Halbach Magnet Based on Magnetic Sheet Arrays.” Journal of Magnetic Resonance 339 (June 2022): 107210.

https://doi.org/10.1016/j.jmr.2022.107210.

Halbach magnet has great potential in nuclear magnetic resonance device, where the field homogeneity requirement puts heavy demands on high efficiency passive shimming (PS) technique. Conventional PS involves a tedious iteration process of the magnetic block/sheet number and positions optimization. In this paper, we propose a PS method based on magnetic sheet arrays (MSAs) targeting at spherical harmonic basis up to the 3rd order including dedicated composition of Y(4Z2-X2-Y2), Z3 and X(4Z2-X2Y2) (n = 3, m = À1,0,1) with cross terms to implement structural field compensation. With this approach, the homogeneity of a 0.5 T Halbach magnet was improved from the original 811 ppm to 4.7 ppm in a L15 mm  R2.5 mm water sample. Rough shimming in another 0.93 T Halbach magnet also improved the homogeneity from 1103 ppm to 125 ppm in R2.5 mm sphere. This work provides a flexible, convenient PS method based on MSAs for compact Halbach magnet, which can be applied in NMR spectrometers and other high homogeneity application circumstances.